Specification
"Main Pump (Flow) - Test (Constant Horsepower Flow Control)"     | ||||||
---|---|---|---|---|---|---|
Date     | Machine Serial Number     | |||||
Machine Settings     | Power Shift Pressure     | Engine Speed     | ||||
1900 kPa (276 psi) |
Engine speed dial "10" AEC switch OFF (1800 rpm)     |
|||||
    | Measuring Points     | |||||
9000 kPa (1305 psi) |
15000 kPa (2175 psi)     |
23000 kPa (3336 psi)     |
30000 kPa (4350 psi)     |
33000 kPa (4786 psi)     |
||
Flow measured liter/min (US gpm)     | Drive Pump     |     |     |     |     |     |
Idler Pump |     |     |     |     |     | |
Oil temperature °C (°F)     |     |     |     |     |     | |
    |     |     |     | |||
Engine speed (rpm)     |     |     |     |     |     | |
    |     |     |     | |||
Flow corrected liter/min (US gpm)     | Drive Pump     |     |     |     |     |     |
Idler Pump |     |     |     |     |     | |
Specification for flow rate liter/min (US gpm)     | New     | 208 ± 5 L/min (55 ± 1.3 US gpm)     |
181 ± 10 L/min (48 ± 2.6 US gpm)     |
139 ± 8 L/min (37 ± 2.1 US gpm)     |
108 ± 6 L/min (29 ± 1.6 US gpm)     |
103 ± 6 L/min (27 ± 1.6 US gpm)     |
Service Limit | 189 L/min (49.9 US gpm)     |
143 L/min (37.8 US gpm)     |
108 L/min (28.5 US gpm)     |
74 L/min (19.5 US gpm)     |
70 L/min (18.5 US gpm)     |
Illustration 1 | g02769436 |
P-Q characteristic curve |
Note: Specifications for output flow rates are based on an engine speed of 1800 rpm. To attain more accurate test results, measured flow should be corrected with the following calculation.
Corrected flow     | =     | Measured flow x 1800 rpm     | |||
measured rpm |
Note: Specifications for output flow rates are based on a power shift pressure at the pump of 1900 kPa (276 psi). The output flow rate changes approximately 6 L/min (1.6 US gpm) for each 100 kPa (14 psi) of power shift pressure in the lower range of the constant horsepower control. The output flow rate changes approximately 3 L/min (0.8 US gpm) for each 100 kPa (14 psi) of power shift pressure in the higher range of the constant horsepower control.
Note: If the output flow rate of the pump is not within the specification, disconnect the negative flow control line from the pump that is being tested and perform the test again.
Note: Flow measurements must be done only during pressure rise.
Introduction
This test measures individual pump flow at the pump. Each pump is tested individually for output flow at specified pressures. The pump that is not being tested is not under load and the pump remains at low-pressure standby.
The following procedures must be completed before performing the test procedure.
- Set the pilot pressure to the specifications in Testing and Adjusting, "Relief Valve (Pilot) - Test and Adjust" for the machine.
- Calibrate the power shift pressure. Refer to Testing and Adjusting, "Solenoid Valve (Proportional Reducing) - Calibrate (Power Shift Pressure)" for the machine.
- Calibrate the proportional reducing valve for negative flow control. Refer to Testing and Adjusting, "Solenoid Valve (Proportional Reducing) - Calibrate (Negative Flow Control)" for the machine.
Required Tools
Illustration 2 | g01623868 |
Tooling (B) 198-4240 Digital Pressure Indicator Gp |
Illustration 3 | g02535237 |
Tooling (C) |
Illustration 4 | g02084974 |
Tooling (D) (Flow meter tool layout) |
Required Tools     | ||||
---|---|---|---|---|
Tool     | Item     | Part Number     |
Description     | Qty     |
B     | 198-4240     | Digital Pressure Indicator Gp     |     | |
    | B1     | 198-4234     | Indicator     | 1     |
B2 | 198-4237     | Pressure Sensor 3445 kPa (500 psi)     |
1     | |
B2 | 198-4239     | Pressure Sensor 41,368 kPa (6,000 psi)     |
2     | |
B3 | 198-4236     | Extension Cable     | 3     | |
C     | 348-5430     | Multi-Tool Gp     |     | |
    | C1     | 285-0908     | Multi-Tool     | 1     |
C2 | 6V-2198     | Cable As     | 1     | |
C3 | 9U-5140     | Photo Pickup     | 1     | |
C4 | 6V-3137     | Base     | 1     | |
C5 | 1U-6605     | Reflective Tape     | 1     | |
C6 | 6V-3138     | Extension     | 1     | |
D     | D1     | 4C-9910     | Portable hydraulic tester (flow meter)     | 1     |
D2 | 6K-6307     | O-Ring Seal     | 2     | |
D3 | 8C-9026     | Adapter     | 2     | |
D4 | 7M-8485     | O-Ring Seal     | 9     | |
D5 | 4C-6482     | Fitting     | 4     | |
D6 | 4C-6481     | Coupler     | 4     | |
D7 | 4C-8767     | Coupling     | 4     | |
D8 | 5P-1010     | Sleeve     | 4     | |
D9 | 5P-0201     | Hose     | 2     | |
D10 | 9S-4191     | O-Ring Plug     | 1     | |
D11 | 3J-1907     | O-Ring Seal     | 1     | |
D12 | 1U-8303     | Fitting     | 1     | |
D13 | 8T-4184     | Bolt     | 4     | |
D14 | 8T-4223     | Hard Washer     | 4     | |
D15 | 6V-0400     | Half Flange     | 2     | |
D16 | 1U-8292     | Adapter     | 1     | |
D17 | 1P-3704     | Rectangular Seal     | 1     | |
D18 | 9U-7335     | Blocking Cover     | 1     | |
D19 | 4I-6141     | Adapter     | 1     | |
D20 | 5K-9090     | O-Ring Seal     | 2     | |
D21 | 6V-9840     | Swivel Tee     | 1     | |
F     | 311-1362     | Vacuum Cap     | 1     |
ReferenceRefer to Tool Operating Manual, NEHS1087 for operating instructions for the 348-5430 Multi-Tool Gp .
Note: Use Software License Card, NETG5049 with part number 348-5430 Multi-Tool Gp .
Machine Preparation
Illustration 5 | g02355508 |
Flow meter tool layout Tooling (D) (4) Delivery line for drive pump (5) Delivery line for idler pump (6) Bolt (7) Washer (8) Half flange (9) Seal (10) Makeup line (11) Seal (12) Elbow |
Drive Pump
NOTICE |
---|
Care must be taken to ensure that fluids are contained during performance of inspection, maintenance, testing, adjusting, and repair of the product. Be prepared to collect the fluid with suitable containers before opening any compartment or disassembling any component containing fluids. Refer to Special Publication, NENG2500, "Dealer Service Tool Catalog" for tools and supplies suitable to collect and contain fluids on Cat products. Dispose of all fluids according to local regulations and mandates. |
Note: Perform the test for the drive pump and the test for the idler pump one at a time.
Illustration 6 | g02646703 |
Pump compartment (1) Pressure tap (delivery pressure for drive pump) (2) Pressure tap (delivery pressure for idler pump) (3) Pressure tap (power shift pressure) (4) Delivery line for drive pump (5) Delivery line for idler pump |
- Position the machine on level ground.
- Stop the engine.
- Release pressure in the hydraulic system. Refer to Testing and Adjusting, "System Pressure - Release".
- Attach 311-1362 Vacuum Cap onto the hydraulic tank. Attach an air supply hose onto the assembly. Apply 276 to 414 kPa (40 to 60 psi). This procedure will pull a vacuum on the hydraulic system.
- Install the following tools in accordance with the flow meter tool layout. Refer to Illustration 5.
- Disconnect delivery line (4) from the drive pump.
- Install seal (9), half flanges (8), and blocking cover (D18) to the end of the delivery line with bolts (6) and washers (7) .
- Assemble and install seal (D17), adapter (D16), seals (D4), fitting (D12), seal (D11), plug (D10), and fitting (D5) on the drive pump by using half flanges (D15), bolts (D13), and washers (D14) .
- Disconnect makeup line (10) from elbow (12) at the swing motor.
Show/hide tableIllustration 7 g02646783
Swing motor (top view)
(10) Makeup line
(11) Seal
(12) Elbow
- Assemble and install seals (D20), tee (D21), adapter (D19), seal (D4), and fitting (D5) to elbow (12) at the swing motor.
- Install seal (11) and makeup line (10) to tee (D21) .
- Remove 311-1362 Vacuum Cap and install the filler plug for the hydraulic tank.
- Connect portable hydraulic tester (D) and test hoses between coupling assembly at the drive pump and coupling assembly at the swing motor.
- Attach a 41,368 kPa (6,000 psi) pressure sensor to pressure tap (1) for delivery pressure of the drive pump. Attach a 3445 kPa (500 psi) pressure sensor to pressure tap (3) for power shift pressure.
- Disconnect connector (13) from proportional reducing valve (15) .
Illustration 8 | g02646796 |
Proportional reducing valves (negative flow control) (13) Connector (14) Connector (15) Proportional reducing valve for "NFC1" (16) Proportional reducing valve for "NFC2" |
- Install tooling (C) on the engine to monitor engine speed.
Idler Pump
NOTICE |
---|
Care must be taken to ensure that fluids are contained during performance of inspection, maintenance, testing, adjusting, and repair of the product. Be prepared to collect the fluid with suitable containers before opening any compartment or disassembling any component containing fluids. Refer to Special Publication, NENG2500, "Dealer Service Tool Catalog" for tools and supplies suitable to collect and contain fluids on Cat products. Dispose of all fluids according to local regulations and mandates. |
Note: Perform the test for the drive pump and the test for the idler pump one at a time.
Illustration 9 | g02646703 |
Pump compartment (1) Pressure tap (delivery pressure for drive pump) (2) Pressure tap (delivery pressure for idler pump) (3) Pressure tap (power shift pressure) (4) Delivery line for drive pump (5) Delivery line for idler pump |
- Position the machine on level ground.
- Stop the engine.
- Release pressure in the hydraulic system. Refer to Testing and Adjusting, "System Pressure - Release".
- Attach 311-1362 Vacuum Cap onto the hydraulic tank. Attach an air supply hose onto the assembly. Apply 276 to 414 kPa (40 to 60 psi). This procedure will pull a vacuum on the hydraulic system.
- Install the following tools in accordance with the flow meter tool layout. Refer to Illustration 5.
- Disconnect delivery line (5) from the idler pump.
- Install seal (9), half flanges (8), and blocking cover (D18) to the end of the delivery line with bolts (6) and washers (7) .
- Assemble and install seal (D17), adapter (D16), seals (D4), fitting (D12), seal (D11), plug (D10), and fitting (D5) on the idler pump by using half flanges (D15), bolts (D13), and washers (D14) .
- Disconnect makeup line (10) from elbow (12) at the swing motor.
Show/hide tableIllustration 10 g02646783
Swing motor (top view)
(10) Makeup line
(11) Seal
(12) Elbow
- Assemble and install seals (D20), tee (D21), adapter (D19), seal (D4), and fitting (D5) to elbow (12) at the swing motor.
- Install seal (11) and makeup line (10) to tee (D21) .
- Remove 311-1362 Vacuum Cap and install the filler plug for the hydraulic tank.
- Connect portable hydraulic tester (D) and test hoses between coupling assembly at the idler pump and coupling assembly at the swing motor.
- Attach a 41,368 kPa (6,000 psi) pressure sensor to pressure tap (2) for delivery pressure of the idler pump. Attach a 3445 kPa (500 psi) pressure sensor to pressure tap (3) for power shift pressure.
- Disconnect connector (14) from proportional reducing valve (16) .
Illustration 11 | g02646796 |
Proportional reducing valves (negative flow control) (13) Connector (14) Connector (15) Proportional reducing valve for "NFC1" (16) Proportional reducing valve for "NFC2" |
- Install tooling (C) on the engine to monitor engine speed.
Test Procedure
Note: Perform the test for the drive pump and the test for the idler pump one at a time.
To prevent personal injury or equipment damage from failure of the hydraulic test equipment or associated circuit components because of blocked pump flow, make sure that the test equipment valves are fully open before starting the engine. |
To prevent personal injury and/or equipment damage from failed lines or components while the hydraulic test equipment is returned to the open flow position, slowly open the hydraulic test equipment valve while monitoring the pump flow. If pump flow does not increase as the valve is opened, shut the engine off and determine what is preventing the pump from upstroking. |
- Start the engine.
- Increase the hydraulic oil temperature to 55° ± 5°C (131° ± 9°F).
- Place the machine controls at the following settings: engine speed dial "10" and AEC switch OFF.
- Enter the service mode.
- Input the password "9992".
- Select "Override Parameters".
- Set the engine speed to 1800 rpm.
Illustration 12 | g02791925 |
- Set the power shift pressure to 1900 kPa (276 psi). Confirm the power shift pressure setting at pressure tap (3) .
- In order to perform flow measurements for either one of the pumps, the other pump must not be under load.
- Turn valve (17) on the portable hydraulic tester clockwise. Record pump flow at each of the following pressures in Table 1. Use the pressure sensor at pressure tap (1) to monitor the delivery pressure of the drive pump or use pressure tap (2) to monitor the delivery pressure of the idler pump.
Illustration 13 | g02599896 |
Portable hydraulic tester (flow meter) (D1) Portable hydraulic tester (flow meter) (17) Valve |